Preparing of Stocks Solutions, Handling and Storage
Opening Vials Containing Lyophilized Products:
Spin down before opening. The pellet may have relocated during shipment.
Hybridization Probes for the LightCycler® Instruments
HybProbes are delivered in fixed molar amounts of 300 pmol, 1 nmol, 3 nmol, or 30 nmol.
Do not freeze lyophilized probes. Products are stable at room temperature. Protect from light exposure.
Usage with the LightCycler® Instruments
Preparation of Stock Solutions, 3 µM for Probes and 10 µM Primers
|
Amount |
300 pmol |
1 nmol |
3 nmol |
30 nmol |
|
HybProbe 3FL or 5LC |
100µl |
333 µl |
1000µl |
1000µl |
|
|
3µM |
3µM |
3 µM |
30µM 1:10 |
|
Concentration |
3µM |
3µM |
3µM |
3µM |
|
Primer iLC |
|
1 nmol |
3 nmol |
30 nmol |
|
|
|
100µl |
300µl |
300µl 1:10 |
|
Concentration |
|
10µM |
10µM |
10µM |
|
PCR-Primer |
1.0 OD260 ~ 5 nmol |
10-15 nmol |
|
|
500µl |
1000-1500µl |
|
Dilution of dissolved primers |
Samples (for a 10µM solution) |
14µM 4µl aq + 10µl oligo |
Storage
Store lyophilized products at room temperature or refrigerated. Do not freeze lyophilized products!
Store dissolved probes refrigerated (4°C) or frozen (-20°C). Avoid multiple cycles of freezing and thawing. We suggest preparing aliquots to be stored frozen. Keep one aliquot refrigerated for daily work. Please read the product description carefully for dilution instructions. Add 100 µl/1 nmol for a 10 µM Stock solution. For dissolved products > 10 µM take a volume = concentration in µM minus 10 µl water and ad 10 µl oligo to prepare a 10 µM solution.